www.szutmedical.com
    HOME >SYRINGE PUMP >SYRINGE PUMP
   
 
SYRINGE PUMP
 
 
 
 
 
Syringe Pump < 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

 

 

 

 

 

 

 

 

 

 

SK-500I/500IA Syringe Pump

Function & Features:
*
Awarded with ISO13485:2003(TUV Germany) and CE Certificated(CE0197)
*
All Standard syringes of 1ml,5ml,10ml,20ml,50ml are compatible.
*
The machine diagnoses the syringe size automatically, and with a simple syringe specification setting, the syringe pump can guarantee an accurate and safe injection.
*
Unique function of bedside injection supervision which makes patients and their relatives as well as nurses more assured.
*
Excellent injection function and operation.
*
Even small-volume injection is accurate.
*
Double CPU system ensures safe and reliable injection process.
*
Driven by precision step motor, no pulsating waves shall occur during even low-rate injection.
*
Accurate&sensitive pressure sensor detects occlusion pressure accurately.
*
Alarms are given in case of malfunction or operation error and the machine stops running automatically.

 

Technical Parameters:
Max Infusion Rate 500ml/h (Different syringes have different max rates).
Infusion Rate 10ml syringe: 0.1-100ml/h
20ml syringe : 0.1-200ml/h
50ml syringe : 0.1-500ml/h
Bolus Rate

10ml syringe : 100ml/h
20ml syringe : 200ml/h
50ml syringe : 500ml/h

Accumulated infusion volume 0.1-999.9ml
Accuracy ¡À3£¥
Power Supply AC110-240V, 50/60Hz, 25W.
Battery Rechargeable lithium battery, 7.4 V, 1650mAh,compliance with the requirements of IEC 60601-1.
Max Power Consumption 25W, running more than 5 hours after being fully charged.
Battery Charge When syringe pump is connected to AC power, the battery will automatically start charging (need8~14hrs).
Fuses F2.0AL, 250V.
Displayed Info Flow rate, accumulated infusion volume, syringe size, battery capacity, bed No., AC power connecting indication
Alarm Info infusion completion, near completion, occlusion, low battery, syringe abnormal, control abnormal, no AC alarm
Overall Size 288¡Á130¡Á122 mm£¨LxWxH£©
Self-Weight £¼2.4kg
Classification Type BF
Waterproof IP21
Electrical Shock Proof Battery: Class II; AC Power: Type BF .
Shell Material ABS plastic
Operation Environment Environment temperature +5¡æ--+40¡æ, atmosphere pressure 86~106kPa£¬relative humidity 20%~90%
Storage Environment Environment temperature -15¡æ- +50¡æ, atmosphere pressure 86~106kPa£¬relative humidity20%~90%
Applicable Syringes 1ml,5ml,,10ml, 20ml and 50ml syringe
Standard

Comply with the standards of EN ISO14971:2000 +A1:2003, IEC60601-1:1995+A1+A2, EN IEC 60601-1-2:2001,IEC60601-2-24,1998, EN 980:2003, EN 1041:1998, EN60601-1-4:1996+A1:1999,
EN60601-1-6:2004.

Operating Mode Continous operation

 

SK-500II Syringe Pump

Function & Features:
*
Awarded with ISO13485:2003(TUV Germany) and CE Certificated(CE0197)
*
Three Kinds of Modes: Rate Mode, Time Mode, Body Weight Mode.
*
Double CPU, make the process of injection safe and more reliable.
*
Accurate&sensitive pressure sensor detects occlusion pressure accurately.
*
Injection time,dose,solution can be checked during the process of injection.
*
Precision step motor.
*
There is no pulsating wave even during low-rate injection.
*
KVO rate can be selected from 0.1ml/h to 5ml/h.
*
Excellent injection function and operation.
*
Even small-volume injection is accurate.
*
All Standard syringes of 5ml, 10ml, 20ml, 30ml and 50ml are compatible.
*
It will self-diagnose and display the syringe size and set parameter automatically in the form of menu.
*
Alarms are given in case of malfunction or operation error and the machine stops running automatically.
*
Save the preset volume automatically when there is unpredictable switch off, ensuring the accuracy of injection volume.
 
Technical Parameters:
Max Infusion Rate
1500ml/h (Different syringes have different max rates).
Infusion Rate
5ml syringe: 0.1-100ml/h
10ml syringe: 0.1-200ml/h
20ml syringe: 0.1-400ml/h
30ml syringe : 0.1-600ml/h
50ml syringe : 0.1-1500ml/h
Bolus Rate

5ml syringe: 100ml/h
10ml syringe: 200ml/h
20ml syringe: 400ml/h
30ml syringe : 400-600ml/h
50ml syringe : 400-1500ml/h

Injection Increment
0.1ml/h
Injection Mode
1. Rate Mode 2. Time Mode 3. Body Weight Mode
Rate Mode
Speed: 0.1-1500ml/h
(Max value depends on the syringe specifications.)
Time Mode
Time:1~2000 minutes
Accumulated injection volume: 0.1-999.9ml
Body Weight Mode
Weight: 0.1~300.0kg
Drug: 0.1~999.9(mg)
Volume: 0.1~999.9ml
Dose: 0.1¡«9999.9
(Restricted to unit,drug,solution,weight and syringe specification.)
Unit: mg/kg/h, ug/kg/min
KVO Rate
Adjustable range: 0.1ml/h¡«5.0ml/h (begin the KVO function under the occlusion situation) ; Pressing STOP key to cancel the KVO function when it is unnecessary in use.
IP Grade
IP21
Injection Pressure
The max pressure is 0.3Mpa, Occlusion alarm pressure value is 40kPa¡«160kPa when pressure value is minimum, injection rate is 25ml/h, the longest alarm time is 10 minutes. When pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes.
Preset Volume
0.1ml-9999.9ml
Preset Time
0~99 hours, 0~59 minutes
Accumulated Infusion Volume
0.1-9999.9ml
Accuracy
¡À3£¥
Power Supply
AC110-240V, 50/60Hz
Battery
Rechargeable lithium battery, 7.4 V, 1600mAh,compliance with the requirements of IEC 60601-1.
Max Power Consumption
25W, running more than 2 hours after being fully charged.
Battery Charge
When syringe pump is connected to AC power, the battery will automatically start charging (need8~14hrs).
Fuses
F2.0AL, 250V.
Displayed Info
Flow rate; Accumulated Injection Volume; Syringe Specification; Battery Capacity; Bed No.; AC Power Indicator, Operation Indication, etc.
Alarm Info
Near Completion, Over, Occlusion, Low Battery, Abnormal 1 (communications failure), Abnormal 2 (pump stuck), Abnormal 3 (wrong parameter), Volume Limited Alarm, No AC Power
Overall Size
288¡Á130¡Á122 mm£¨LxWxH£©
Self-Weight
£¼3.0kg
Classification
Class II, Type BF
Electrical Shock Proof
Battery: Class II; AC Power: Type BF .
Shell Material
ABS plastic
Operation Environment
Environment temperature +5¡æ--+40¡æ, atmosphere pressure 86~106kPa£¬relative humidity 20%~80%
Storage Environment
Environment temperature -15¡æ- +50¡æ, atmosphere pressure 50~106kPa£¬relative humidity20%~95%
Applicable Syringes
5ml,,10ml, 20ml,30ml and 50ml syringe
Standard

Comply with the standards of EN ISO14971:2000 +A1:2003, IEC60601-1:1995+A1+A2, EN IEC 60601-1-2:2001,IEC60601-2-24,1998, EN 980:2003, EN 1041:1998, EN60601-1-4:1996+A1:1999,
EN60601-1-6:2004.

Electrical Safety
In accordance with the stipulation of GB9706.1-1995

 

SK-500III Syringe Pump(Double-channels)

Function & Features:
*
Awarded with ISO13485:2003(TUV Germany) and CE Certificated(CE0197)
*
All Standard syringes of 1ml,5ml,10ml,20ml,50ml are compatible.
*
Multi-channels (Double-channels), sub-rate injection.
*
Four kinds of injection modes: Rate Mode,Time mode,Body Weight Mode,Link Mode.
*
Innovative design of outer appearance gives a strong sense of science and technology.
*
Double CPU makes the process of injection safer and more reliable.
*
Brand-new STM32 ARM-Cortex M3 platform, optimized circuit design, operation rate more faster and stable.
*
The use of TFT color LCD makes the screen display more remarkable.
*
Digital menu, easy operation; functions fully meet the clinical demand.
*
Ample and user-friendly indicator information; accurate and complete alarm information to ensure safe injection.
*
Aluminum alloy base, integrative pump device makes more smooth operation and easier installation and maintenance.
 
Technical Parameters:
Max Infusion Rate
1500ml/h (Different syringes have different max rates).
Infusion Rate
5ml syringe: 0.1-100ml/h
10ml syringe: 0.1-200ml/h
20ml syringe: 0.1-400ml/h
30ml syringe : 0.1-600ml/h
50ml syringe : 0.1-1500ml/h
Bolus Rate

5ml syringe: 100ml/h
10ml syringe: 200ml/h
20ml syringe: 400ml/h
30ml syringe : 400-600ml/h
50ml syringe : 400-1500ml/h

Injection Increment
0.1ml/h
Injection Mode
1. Rate Mode 2. Time Mode 3. Body Weight Mode 4. Link Mode
Rate Mode
Speed: 0.1-1500ml/h
(Max value depends on the syringe specifications.)
Time Mode
Time:1~2000 minutes
Accumulated injection volume: 0.1-999.9ml
Body Weight Mode
Weight: 0.1~300.0kg
Drug: 0.1~999.9(mg)
Volume: 0.1~999.9ml
Dose: 0.1¡«1500.00mg/kg/h
(Restricted to unit,drug,solution,weight and syringe specification.)
Unit: mg/kg/h, ug/kg/min
KVO Rate
Adjustable range: 0.1ml/h¡«5.0ml/h (begin the KVO function under the occlusion situation) ; Pressing "0" key to cancel the KVO function when it is unnecessary in use.
Injection Pressure
The max pressure is 0.3Mpa, Occlusion alarm pressure value is 40kPa¡«160kPa when pressure value is minimum, injection rate is 25ml/h, the longest alarm time is 10 minutes. When pressure value is minimum, injection rate is 5ml/h, the longest alarm time is 10 minutes.
Preset Volume
0.1ml-9999.9ml
Preset Time
0~99 hours, 0~59 minutes
Accumulated Infusion Volume
0.1-9999.9ml
Accuracy
¡À3£¥
Power Supply
AC110-240V, 50/60Hz
Battery
Rechargeable lithium battery, 7.4 V, 1600mAh,compliance with the requirements of IEC 60601-1.
Max Power Consumption
25W, running more than 2 hours after being fully charged.
Battery Charge
When syringe pump is connected to AC power, the battery will automatically start charging (need8~14hrs).
Fuses
F2.0AL, 250V.
Displayed Info
Flow rate; Accumulated Injection Volume; Syringe Specification; Battery Capacity; Bed No.; AC Power Indicator, Operation Indication, etc.
Alarm Info
Near Completion, Over, Occlusion, Low Battery, Abnormal 1 (communications failure), Abnormal 2 (pump stuck), Abnormal 3 (wrong parameter), Volume Limited Alarm, No AC Power
Overall Size
320¡Á200¡Á110 mm£¨LxWxH£©
Self-Weight
£¼3.0kg
Classification
Class II, Type BF
Electrical Shock Proof
Battery: Class II; AC Power: Type BF .
Shell Material
ABS plastic
Operation Environment
Environment temperature +5¡æ--+40¡æ, atmosphere pressure 86~106kPa£¬relative humidity 20%~80%
Storage Environment
Environment temperature -15¡æ- +50¡æ, atmosphere pressure 50~106kPa£¬relative humidity20%~95%
Applicable Syringes
5ml,,10ml, 20ml,30ml and 50ml syringe
Standard

Comply with the standards of EN ISO14971:2000 +A1:2003, IEC60601-1:1995+A1+A2, EN IEC 60601-1-2:2001,IEC60601-2-24,1998, EN 980:2003, EN 1041:1998, EN60601-1-4:1996+A1:1999,
EN60601-1-6:2004.

Electrical Safety
In accordance with the stipulation of GB9706.1-1995

 

   
  Suzhou UnionTech Imp. & Exp. Co. Ltd
Add£ºRoom 2302, Jiafu International Plaza, No.515, Western Ganjiang Road, Suzhou, China P.C. 215004
Tel£º0086-512-68632895  69325080 Fax£º0086-512-68632896    Mobile£º13706219549
Email£ºsales@szutmedical.com yangyijun@szuniontech.com